octave-bug-tracker
[Top][All Lists]
Advanced

[Date Prev][Date Next][Thread Prev][Thread Next][Date Index][Thread Index]

[Octave-bug-tracker] [bug #58252] [octave forge] (instrument-control) us


From: Walter Sjursen
Subject: [Octave-bug-tracker] [bug #58252] [octave forge] (instrument-control) ustmc_read() does not read all the waveform data from oscilloscope
Date: Thu, 30 Apr 2020 12:58:39 -0400 (EDT)
User-agent: Mozilla/5.0 (X11; Ubuntu; Linux x86_64; rv:75.0) Gecko/20100101 Firefox/75.0

Follow-up Comment #7, bug #58252 (project octave):

I did not try the test version of instrument-control yet. I did run a test
where I increased the number of waveform points requested and plot the number
of returned points vs. request points (see the attached file "figure1.png").
At about 280 waveform points (3930 bytes of return data + header), it starts
falling apart. Below 280 waveform points, it work's fine.

(file #48976)
    _______________________________________________________

Additional Item Attachment:

File name: figure1.png                    Size:45 KB
    <https://savannah.gnu.org/file/figure1.png?file_id=48976>



    _______________________________________________________

Reply to this item at:

  <https://savannah.gnu.org/bugs/?58252>

_______________________________________________
  Message sent via Savannah
  https://savannah.gnu.org/




reply via email to