From: | Walter Sjursen |
Subject: | [Octave-bug-tracker] [bug #58252] [octave forge] (instrument-control) ustmc_read() does not read all the waveform data from oscilloscope |
Date: | Thu, 30 Apr 2020 12:58:39 -0400 (EDT) |
User-agent: | Mozilla/5.0 (X11; Ubuntu; Linux x86_64; rv:75.0) Gecko/20100101 Firefox/75.0 |
Follow-up Comment #7, bug #58252 (project octave): I did not try the test version of instrument-control yet. I did run a test where I increased the number of waveform points requested and plot the number of returned points vs. request points (see the attached file "figure1.png"). At about 280 waveform points (3930 bytes of return data + header), it starts falling apart. Below 280 waveform points, it work's fine. (file #48976) _______________________________________________________ Additional Item Attachment: File name: figure1.png Size:45 KB <https://savannah.gnu.org/file/figure1.png?file_id=48976> _______________________________________________________ Reply to this item at: <https://savannah.gnu.org/bugs/?58252> _______________________________________________ Message sent via Savannah https://savannah.gnu.org/
[Prev in Thread] | Current Thread | [Next in Thread] |